Contact us to discuss your application / options and pricing.

    SKU: VTT10-MH Category: Tags: , , ,

    Description

    Specifications

    Features & Benefits

    • Accuracy: Temperature: According Tables 6.2, 6.3 and 6.4
      4-20 mA Outputs: ± 0.1% of Calibrated Span
    • HART Power Supply (PWR): 12 to 45 Vdc / 4-20 mA according to NAMUR-NE43
    • Open-Collector Power Supply (IN): 24 Vdc ±5%
    • Communication Protocol HART® 7
    • Classified Areas: Explosion Proof (with certified enclosure) and Intrinsically Safe (pending)
    • Ambient Temperature Limits: -20 to 70°C
    • Ambient Temperature Effect (for 1°C variation):
      • Resistive Sensors: ± 0.0052% of Ohm reading
      • Millivoltage Sensors: ± 0.001% of mV reading
    • Reading Stability: ±0.1% of reading or 0.1°C – the highest value
      RTD: 3 years; Thermocouples: 2 years
    • Configuration: Local, EDDL, FDT/DTM and Android® Tools
    • Mounting DIN Rail or in the field with explosion proof enclosure
    • Protection Degree: IP20 or IP65 (with certified enclosure)
    • Indication: 5-digit, rotative, multifunctional LCD display
    • Housing Material: Aluminum / Plastic
    • Approximated Weight: 540 g (without certified enclosure)