Compliance West’s line of Leakage Current Networks are built in accordance with IECEE OD-5013, “Leakage (Touch) Current Measurement Instruments”; and are in compliance with IECEE OD-5014, “Instrument Accuracy Limits, specifically “Leakage (Touch) Current” Section; and IEC 60990:2016.  Their Leakage Current Networks use very accurate components which do not drift due to heat or age, and they specially trim each component in each leakage current box they make to be accurate.  Compliance West tests the devices to be accurate to 5% from 20Hz to 1MHz, and use metal enclosures to minimize interference.

Designed to provide a fast, safe and reliable method of measuring unweighted touch current, touch current weighted for perception or reaction and touch current weighted for let-go. Measures DC and AC currents: calibrated frequency response 20Hz to 1MHz (Configurations 1, 2, 3, 4 and 6). Designed and built in accordance with IECEE OD-5013 and IECEE OD-5014, as well as in accordance with the applicable transfer impedance requirements in IEC 60990.

Specifications

Features & Benefits

  • Portable, lightweight
  • Measures 4.5 x 1.2 x 2.4 in.
  • No batteries or power required
  • Custom test leads available
  • Provided with NIST calibration; A2LA Calibration available; 20 Hz-1Mhz
  • One year warranty